CARDIOVASCULAR SYSTEMS

Oxygenator Systems

gish biomedical header
» Oxygenator System Specs  » Oxygenator Holder  » Oxygenator Air Challenge

Vision™ Oxygenator Air Challenge Study


In an independent laboratory test, the Vision® Oxygenator System was compared to equivalent membranes in their ability to handle gaseous emboli.  Each device was placed into a circuit where a proprietary digital imaging system was utilized to optically assess gaseous emboli activity at the outlet of the oxygenators.  Utilizing a centrifugal pump to generate clinically significant flow rates, boluses of air were injected into the venous inlet of the oxygenators.  Monitoring the arterial side of the device in one-quarter millisecond intervals (.25 ms), the digital imaging system recorded emboli exiting the oxygenators.  The digital imaging system was programmed to isolate and identify the various micron sizes appearing at the outlet.  The results show emboli ranging from 20 micron to 120 micron passing through many devices, while the Vision Oxygenator System virtually eliminated the undesired air.




Studies, White Papers and Presentations
The Proposed Use of Performance Indices to Evaluate and Compare the "GISH Vision" Membrane Oxygenator (590kb pdf)
Effects of Surface-Modified Hollow-Fiber Membrane Oxygenators on Circulating Platelet Counts During Cardiopulmonary Bypass (3.23mb pdf)
   


GISH 2009 online catalog GISH Online Catalog